International Peer-Reviewed Journal•Open Access•ISSN 2456-8880
irejournals@gmail.com•+91-7433024337

Home / Current Issue / Paper 1707535

1707535 Vol 8 · Issue 9 Download Paper

Assessment of Entrance Skin Dose of X-ray Using Calculation Techniques.

Aminu Dahiru Prof. Felix Wilfred Burari Nasir Muhammad Umar Ali Ado Umarfarouq Dahiru Ahmad Abdulkadir Hassan Muhammad Bashir Usama Yusuf

Subject area: Biological & Medical Sciences  ·  Area of research: Medical Physics

Abstract

skin dose estimation is usefull and well established method for quality control in diagonostic radiology and provide valuable information about the efficacy of the department. This study assessed. skin dose estimation using Mathematical Formula at Yobe State University Teaching Hospital Damaturu, Yobe State Nigeria. The main objective of the study was to calculate entrance skin dose of X-ray using Mathematical formula and to compare the result with the same measured dose using survey meter to test the percentage accuracy of the formula. The study employed the use of prospective research method. 50 patient?s datas were analysed and their ESD is obtained both mathematically and physically using survey meter and the two results are in good positive correlation using inferential statistical tools, these confirmed the accuracy of the formulas. Findings of the study revealed that out of three formulas used to calculate ESD (Chan & Tsai, Edmond and Kumar) the Kumar method has the strongest positive correlation with the survey meter value in a very close range results. The result further showed that the high population to receive the dose during x-ray examination are obese due to the increase in exposure factor (kVp & mAs) in a high body thickness liable of high attenuation capacity than the small body thickness. Furthermore the contributing factor involved in dose deposition in patients is the faults of the radiation workers dissipating x-ray, (x-ray technician and radiographers) due to the misuse of the machine parameters (exposure factors) and machine manipulation(misuse of beam limiting and centering devices protecting scatter, collimators delineation and beam diaphragm) misuse of anti-scatter grid, etc. The study recommends the training and re-training of radiographers should be focused on, so that issue of dose deposition in patients will be minimized.

Keywords

skin dose, formula, x-ray.

References

[1] Abdulsalam, A. K. (2017): Assessment of dose for conventional X-ray examination at Federal Teaching Hospital Gombe and Specialist Hospitals Gombe, Gombe State. Unpublished master’s thesis, Department of Physics, Gombe State University, Gombe, Nigeria.

[2] Ademola, A. k. Obed, R. I. Adejumobi, C. A. Abidunmi, O. P. Alabi, O. F. & Oladapo, M. (2013). Assessment of Entrance Skin Dose in routine X-ray examination of conventional, skull, abdomen and pelvis of children in five selected hospital in Nigeria. IOSR Journal of applied science 5(2) pp47-50.

[3] Akbar, A. Ehsan, M. Mahboubeh, M. Morteza, S. & Mehran, M. (2015): Measurement of Entrance Skin Dose and Calculation of effective Dose for common Diagnostic X-ray Examinations in Kashan, Iran. Global Journal of Health Science; Vol 7, No. 5 pp202- 205

[4] Bahreyni, M. T. Nazery, M. & Zare, H. (2019): Application of dose-area product meter to measure organ and effective dose in diagnostic radiology at two hospitals, Mashhad, Iran V6 pp 101-113. Available at: http://www.irpall.com/new/pdfs/4b10.pdf.

[5] Bock, R. K. and Vasilescu, A. (2017): The particle detector brief book. Access November 15, 2017 from: http:// www. bonner. rice.edu/~ liuj /manual/ particle- detector brief book/node154.html,

[6] Dietze, G. Eckerman, K. Menzel, H. Stather, J. and Streffer, C. (2015): Basis for dosimetry quantities used in radiological protection ICRP Oxon 19, available at: www.icrp.org/dosimetry.pdf.

[7] Fawaz, H. A. (2017): Assessment of entrance skin dose and effective dose of some routine x- ray examination using calculation techniques. Unpublished Master’s Thesis, King Saud University.

[8] Hart, D. Hillier, M. C. and Wall. B. F. (2016): Doses to patients from medical

[9] X-ray examinations inthe UK, 2016, NRPB Report 279 (Didcot, Oxon: National RadiologicalProtectionBoard). availableat: http://www.hpa.org.uk/radiation/publication/wseriesreports/2014/nrpbw14.pdf

[10] Ibrahim, U. Daniel, I. H. Ayaninola, O. Ibrahim, A. Hamza, A. M. and Umar, A. M. (2014): Determination of Entrance Skin Dose from Diagnostic X-Ray of Human Conventional at Federal Medical Centre Keffi, Nigeria Science World Journal Vol 9 (No 1) 2014 ISSN pp 1597-6343.

[11] International Commission on Radiological Protection, Draft of ICRP Committee 3 (2015): on Diagnostic reference levels in medical imaging.

[12] Ishiguchi, T. wanami, S. I. Kawatsu, S. Ishigaki, T. and Koga S. (2016): Radiation Exposure by Routine Radiographic Examinations, Multicenter Study in Japan with Thermoluminescence Dosimetry and Estimation from the Radiographic Data.

[13] Jes, D. R. online (2017): internet website. Issue 4. Accessed October 10, 2017 from: http://www.place.dawsoncollege.qc.ca/~drjes/issues/4/xrays.html

[14] Kepler, K. Fillipova, I. Lintrop, M. and Servomaa, A. (2014): Pediatric Patient Dosimetry as part of Quality Assurance Programme in Estonia. pp 1-4.

[15] Lampinen, J.S. (2016): Calculating patient specific doses in X-ray diagnostics and fromradio pharmaceuticals. Unpublished master’s Thesis, University of Helsinki.

[16] Li L. B. Wang, J.P. et al., (2001): Medical radiation usage and exposures from medical X-ray diagnosis in Shadong providence of China, Radiation protection dosimetry, Vol. 93pp 261-266.

[17] National Radiation Laboratory, New Zeland (NRLNZ) (2017): XDOSE Computer program

[18] Olowookere, C. J. Babalola I. A. Olayiwola1, M. O. Odina1, G. Obed, R. I. and Bello, T. O. (2019). Comparison of Five Formulas for Assessing Patient Dose from Radiological Examinations Afr J Med Phy, Biomed Eng & Sc, 2019. V 1 pp21-29.

[19] Radiation and Nuclear Safety Authority, Finland (RNSAF) (2018): PCXMC. A PC-based Monte Carlo program for calculation doses in medical X-ray examinations.

[20] Sartin, B. (2018): Radiation therapy, Accessed January 2, 2018, from: http://homepage.mac.com/billsart/radiationsemester3/ahs134lecture17.htm.

[21] Sharifat, I. and Oyeleke, I. O. (2019). Patient entrance skin dose at Minna and Ibadan for common diagnostic radiology examination. Bayero journal of pure and applied science 2(1) pp1-5.

[22] Sprawls, J.P., (2017): Physical Principles of medical imaging (2nd Ed.). (Gaithersburg, Maryland: Aspen publishers Inc.).

[23] Stoner,R.E.(2017):BGSU department of physics and astronomy .Accessed December7, 2017 from: http://www. physics. bgsu. edu/~ stoner/P202 /quantum2 /sld001.htm.

[24] Sulaiman, I. I. and Elsiekh, H. A. (2018): Radiation Doses from some common Paediatric X- ray Examination in Sudan. Rad. Prot. Dosim. 132 pp 64-72.

How to cite this paper

Aminu Dahiru, Prof. Felix Wilfred Burari, Nasir Muhammad Umar; Ali Ado, Umarfarouq Dahiru; Ahmad Abdulkadir, Hassan Muhammad Bashir; Usama Yusuf "Assessment of Entrance Skin Dose of X-ray Using Calculation Techniques." Iconic Research And Engineering Journals Volume 8 Issue 9 2025 Page 886-889
Aminu Dahiru, Prof. Felix Wilfred Burari, Nasir Muhammad Umar; Ali Ado, Umarfarouq Dahiru; Ahmad Abdulkadir, Hassan Muhammad Bashir; Usama Yusuf "Assessment of Entrance Skin Dose of X-ray Using Calculation Techniques." Iconic Research And Engineering Journals, vol. 8, no. 9, Mar. 2025
Aminu Dahiru, Prof. Felix Wilfred Burari, Nasir Muhammad Umar; Ali Ado, Umarfarouq Dahiru; Ahmad Abdulkadir, Hassan Muhammad Bashir; Usama Yusuf (2025). Assessment of Entrance Skin Dose of X-ray Using Calculation Techniques.. Iconic Research And Engineering Journals, 8(9).
Aminu Dahiru, Prof. Felix Wilfred Burari, Nasir Muhammad Umar; Ali Ado, Umarfarouq Dahiru; Ahmad Abdulkadir, Hassan Muhammad Bashir; Usama Yusuf "Assessment of Entrance Skin Dose of X-ray Using Calculation Techniques." Iconic Research And Engineering Journals, vol. 8, no. 9, Mar. 2025.
@article{1707535,
      author = {Aminu Dahiru, Prof. Felix Wilfred Burari, Nasir Muhammad Umar; Ali Ado, Umarfarouq Dahiru; Ahmad Abdulkadir, Hassan Muhammad Bashir; Usama Yusuf},
      title = {Assessment of Entrance Skin Dose of X-ray Using Calculation Techniques.},
      journal = {Iconic Research And Engineering Journals},
      year = {2025},
      volume = {8},
      number = {9},
      pages = {886-889},
      issn = {2456-8880},
      url = {https://www.irejournals.com/formatedpaper/1707535.pdf},
      abstract = {skin dose estimation is usefull and well established method for quality control in diagonostic radiology and provide valuable information about the efficacy of the department. This study assessed. skin dose estimation using Mathematical Formula at Yobe State University Teaching Hospital Damaturu, Yobe State Nigeria. The main objective of the study was to calculate entrance skin dose of X-ray using Mathematical formula and to compare the result with the same measured dose using survey meter to test the percentage accuracy of the formula. The study employed the use of prospective research method. 50 patient?s datas were analysed and their ESD is obtained both mathematically and physically using survey meter and the two results are in good positive correlation using inferential statistical tools, these confirmed the accuracy of the formulas. Findings of the study revealed that out of three formulas used to calculate ESD (Chan & Tsai, Edmond and Kumar) the Kumar method has the strongest positive correlation with the survey meter value in a very close range results. The result further showed that the high population to receive the dose during x-ray examination are obese due to the increase in exposure factor (kVp & mAs) in a high body thickness liable of high attenuation capacity than the small body thickness. Furthermore the contributing factor involved in dose deposition in patients is the faults of the radiation workers dissipating x-ray, (x-ray technician and radiographers) due to the misuse of the machine parameters (exposure factors) and machine manipulation(misuse of beam limiting and centering devices protecting scatter, collimators delineation and beam diaphragm) misuse of anti-scatter grid, etc. The study recommends the training and re-training of radiographers should be focused on, so that issue of dose deposition in patients will be minimized.},
      keywords = {skin dose, formula, x-ray.},
      month = {March},
  }